
Druck ADTS 505 User Manual
5 - 13
K0260 Issue No. 8
5
Controller Stability
This section verifies the control stability.
(1)
Press
LEAK MEASURE/CONTROL
to turn the pressure controllers on.
(2)
Enter an ALT/Ps Aim of 510 mbar with a rate of change of 204 mbar/min.
(3)
Enter a Speed/Qc Aim of 0 mbar with a rate of change of 204 mbar/min.
(4)
Press
ALT/Ps
and wait for the aim values to be achieved.
(5)
Observe the measured values of Ps and Qc on the display for 1 min.
(6)
Check that the displayed value of Ps remains within ±0.068 mbar.
(7)
Check that the displayed value of Qc remains within ±0.068 mbar.
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...