6 - 24
Reference and Specification
K0260 Issue No. 8
Manual Mode
F
IGURE
6-8 M
ANUAL
CONTROL
SELECTIONS
Selecting Control and Leak/Measure Modes
Manual test control always starts in leak measure mode displaying pressures in the last selected units of
measurement. The system can be switched between aero units and pressure units, these units can only be
changed in set-up mode.
Selecting control/leak measure switches between leak measure mode and control mode. In control mode the
set of limits provide protection for the aircraft system and can only be changed in set-up mode.