Druck ADTS 505 User Manual
6 - 17
K0260 Issue No. 8
6
6.3.2 Set-up 2 of 2
[Operational Hours]
•
This display shows the total number of hours for system
operation and the total number of hours of pump
operation. The system records, as a background task,
the total running hours of the unit. At an appropriate
time before the 1000 and 3000 servicings are due a
message is added to the power-up display sequence
reminding the user of the hours to go before a
servicing.
[Software Versions]
•
This display shows the master software version
reference for the product followed by the component
versions used by the display and key-pad [LDK], the system controller and the optional hand
terminal.
[Security]
•
Two PIN codes (level 1 and 2) protect access into two sensitive settings menus. Level 1 allows
access to the sets of operating limits. Level 2 allows access to the system calibration/
characterisation menu. This security menu allows the code or number of digits to be
changed. Both PIN code can be up to 10 digits. A single zero disables the PIN protection,
allowing unrestricted access to the settings.
[Maintain Calibration]
[Test Keypad]
The display shows a schematic of the key-pad with indications of each key press. A test
procedure for the key-pad test is detailed in section 5, Testing and Fault Finding.
[View Manifold]
The display shows a table of the current system data values with real time indications of the
status of the valves and the transducers. This facility provides a method of performance
assessment by showing actual system values which can be compared with the known
system parameters. A fault finding procedure for the system is detailed in section 5, Testing
and Fault Finding.
[Enter PIN]
After entering the correct level 2 PIN, this facility allows access for servicing technicians to
characterise the valves or calibrate the pressure sensors though additional menus.
F1
Setup (2 of 2)
Setup
Operational
Hours
Software
Versions
Security
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...