Druck ADTS 505 User Manual
6 - 25
K0260 Issue No. 8
6
E
XAMPLE
R
ATE
T
IMER
S
CREEN
Rate Timer
Selected in Leak/Measure Mode
1.
Use the touch screen keys to select wait and measure times and use the numeric keys to enter the
values.
Note:
Times entered as whole numbers are interpreted as seconds. Numbers higher than 60 are interpreted
and displayed as minutes and seconds (i.e. 90 = 1 minute 30 seconds = 1:30).
2.
After selecting the timer values and with the required the Ps1, Ps2 and Pt readings, press the START
key. The time counter shows the progress of the test: wait time, then test time. The results, at the end
of the test, shows the timed rates for each channel indicated by “T” after the digits.
3.
Press BACK to return to Leak Measure Mode.
Measure
time
Wait time
Alt and CAS
readings/results