6 - 30
Reference and Specification
K0260 Issue No. 8
F
IGURE
6-10 T
EST
S
EQUENCE
SELECTIONS
Selecting a Test Sequence
1.
From the main menu, select the Custom Test Sequence File. The screen changes to the selected test
screen with the data table from the file.
2.
Use the touch screen keys to select each test procedure value.
3.
After selecting the test procedure value use the
Control/Leak Measure
key to select CONTROL.
4.
The screen changes to CONTROL MODE and shows the actual measurement and rate.
Measure/
Control
Actual rate of
pressure
Actual
channel
readings
Set of aim
values to be
used in control
mode
Only available
in Control Mode
changes to Rate Timer
in Leak/Measure Mode
Rate aim to be
used in this TP
step.
Title
Test
procedure
step