6 - 28
Reference and Specification
K0260 Issue No. 8
2.
Use the Up and Down keys to highlight the required aeronautical and pressure units.
3.
Press SAVE SETTINGS to store in non-volatile memory or press MAIN MENU to store in volatile memory
and return to the Main Menu Screen. Press RESTORE DEFAULTS to restore the last set of stored values
and unit selection saved as a default.
Select limits
Enter the PIN to access the SETUP LIMITS
display. Select one of the preset limits of MAX,
STANDARD and CIVIL or define and enable USER
1 limits. An eight character name/code can
replace USER 1 title.
Default limits
This set of limits, identified as: CIVIL are enabled
when the ADTS 505 is switched on:
MIN ALT
-1000 ft
MAX ALT
50,000 ft
MIN CAS
0.0 knots
MAX CAS
450 knots
MAX MACH
1.0 Mach
MAX ROC
6,000 ft/min
ARINC LIMITS
OFF
ALT CORRECTION
0 ft
MIN Ps
115.972 mbar (2.7 inHg)
MAX Ps
1050.406 mbar (38.1000 inHg)
MIN Qc
0.0 mbar
MAX Qc
368.01 mbar (10.8673 inHg)
MAX Rate Ps 109.85 mbar/min (10.8127 inHg/min)
MAX Rate Qc 109.85 mbar/min (10.8127 inHg/min)
Содержание Druck ADTS 505
Страница 1: ...GE Measurement Control Druck ADTS 505 Air Data Test Set User Manual K0260 ...
Страница 14: ...xii K0260 Issue No 8 Druck ADTS 505 User Manual ...
Страница 54: ...3 26 Operation K0260 Issue No 8 ...
Страница 55: ...Druck ADTS 505 User Manual 3 27 K0260 Issue No 8 3 ...
Страница 56: ...3 28 Operation K0260 Issue No 8 ...
Страница 66: ...5 4 Testing and Fault Finding K0260 Issue No 8 FIGURE 5 1 FAULT FINDING CHART ...
Страница 80: ...5 18 Testing and Fault Finding K0260 Issue No 8 Intentionally left blank ...
Страница 124: ...6 44 Reference and Specification K0260 Issue No 8 Intentionally left blank ...