System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
48
GROUND CABLE
RS85 TUBE STAND
4.1
GND STUD
Ground Plug
MPH A1 MODULE
(AC Distribution Panel)
TOWER
MIS 27302
X1−L
X1−N
PE1−GND
Terminals Strip
2165548
RS85 Power Supply CABLE
MPH 50/65/80 CABINET
MPH A1 MODULE
(AC Distribution Panel)
F15−2
F14−2
Ground Plug
0Vac
230Vac
Gnd
7.1
2140362
MIS 27238
GROUND CABLE
8.1
GND STUD
Ground Stud
GND STUD
TOWER
2183955
MIS 27305
RS85 TUBE STAND
MAXIRAY100 XRAY TUBE
I / O PANEL
2140362
SG 60/100 GROUND CABLE
TS2−GND
3.1
6.1
COMPAX 40E TABLE
SG 60/100 VERTICAL BUCKY
STAND
5.1
(2105836) SUPPLIED with PRINTER−X
PRINTER−X Ground CABLE
PRINTER−X
1.1
GROUND
MPH 50/65/80 CABINET
MAIN POWER
TS−GND
2138905
LINE SAFETY SWITCH
POWER SUPPLY CABLE
MPH A1 A1 MODULE
(AC Distribution Panel)
2.1
CONSOLE GROUND CABLE
Ground Plug
GND (BAR GROUND)
(PU3) MADRID CONSOLE
45553701
MPH A1 MODULE
(AC Distribution Panel)
I / O PANEL
2140362
TABLE Ground CABLE
Ground Plug
TB1 (Ground)
MPH A1 MODULE
(AC Distribution Panel)
Ground Plug
MPH A1 MODULE
(AC Distribution Panel)
Ground Plug
MPH A1 MODULE
(AC Distribution Panel)
GND STUD
MPH A9
MODULE
MIS 27235
MIS 27300
MIS 27301
W19
CABLE #
(45553701)
Содержание Compax 40CE