System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
31
Step
Description
Done
Supporting Documents
ÍÍ
ÍÍ
From previous page
Note:
GRN wires must be connected to
metal covers when available.
See SM COMPAX 40E CE, Dir. 2150299
−
100,
Chapter 1, Installation, Section 7.
MP
H
XT Suspension
RS
−
85
ÍÍ
ÍÍ
MPH 50/65/80: See SM 2137354
−
100,
Job Card IST 002
2 People
2 Hours
COVER INSTALLATION
Compax 40E
SG60
−
100
Tomo
−
Link
See SM SG60 Dir 2150253
−
100 Chapter 1
See SM XT RADIOGRAPHIC SUSPENSION
Dir 2118733
−
100; Chapter 3 or 4, Installation.
See SM SG100 CE Dir 2150297
−
100 Chapter 1
A41
Continues on next page
HHS TEST (If Required)
Test
OK
Yes
No
Step A28
Note:
HHS test or any local, in
−
house,
regional or governmental acceptance tests.
NOTE:
If your country requires the product
rating plate label to be in the language of that
country, follow the instructions on document
#45474336. Paste the appropriate language
label over the English rating plate. The rating
plate labels are: 2379361 (Spanish),
2379361
−
2 (German), 2379361
−
3 (French),
2379361
−
4 (Italian).
1 Person
4 Hours
Refer to Inspection Report 2179651 TDR,
sections 4.12 (40E) and 6.5 (SG60/100)
Refer to Inspection Report 2172166 TDR,
sections 4.13 (40E) and 5.5 (SG60/100)
Refer to Inspection Report 2167647 TDR,
4.12 (40E) and 5.5 (SG60/100)
Содержание Compax 40CE