SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
109
ILLUSTRATION 7
−
5
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR AND RS85 TUBE STAND WITH COMPAX 40E
−
MANUAL AND SG100)
SIGNAL POINT GND
PROTECTIVE
EARTH TERMINAL
REFERENCE TERMINAL
EQUIPOTENTIAL TERMINAL
MIS RUN − FIELD
SUPPLIED WIRE
SAFETY GROUND WIRE
SIGNAL GROUND WIRE
SAFETY & SIGNAL
GROUND WIRE
MIS RUN − CABLE
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE
C
B
A
S
CHASSIS
000
LEGENDE
000
MAIN GROUND
TB1−GND
I/O PANEL GND
COMPAX 40E TABLE
2140362 ( TABLE Ground Cable )
2138905 ( Power Supply Cable )
2140362
2140362 ( SG Ground Cable )
SG 100 WALL STAND
MPH A7
Control Console
MPH A9
Printer−X
MPH A8
Program−X
I +
I −
MPH A3 A2
Main Power
MPH A3 A3
Inverter Board
MPH A3
A1
DC Power
MPH
A1
MPH A2
HV TANK
MPH A5
Real Electronic Panel
MPH 50/65/80 CABINET
MPH A1 A2
Stator Connector
Panel
MPH A6 A4
Wall Bucky
MPH A6 A5
Table Tomo
Collimator
MPH A6 A3
Distribution
Board
MPH A6 A6
II Sensor Board
MPH A6 A8
Plugs Panel
MPH A3
FL1
MPH A4
Front Electronic Panel
GND STUD
Ground Plug
GND STUD
2105836
Printer Ground Cable
(Supplied with Printer)
45553701
(Console Ground Cable)
Note:
All grounding cables are yellow−green.
GND
RS85 TOWER
GND
POWER SUPPLY
2165548
POWER SUPPLY CABLE
RS85 TUBE STAND
X−RAY TUBE
+
SA/SM Ultranet
COLLIMATOR
27301
27300
27302
RS85 Ground Cable
RS 85
24VDC
TUBESTA
ND
POWER
SUPPLY
2165548
POWER SUPPLY CABLE
OPTIONAL
27235
1.1
27238
7.1
7.1
6.1
2.1
3.1
4.1
5.1
2183955
8.1
27305
GND X−RAY TUBE
REFER TO JOBCARD IST002
IN APPROPRIATE VERSION
MPH SERVICE MANUALS
FOR GROUND PLUG LAYOUT
MPH V1 SM:
2137354
−
100
MPH V2/V3 SM: 2165156
−
100
Содержание Compax 40CE
Страница 138: ......