System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
49
1-4-4
Mis Chart
System MPH 50/65/80 with Compax 40E / SG 60 or SG100 / XT Suspension with Tomo Link
Note:
Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:
−
Numbers on the left presented in grey squares are HLA numbers (CABLES #).
−
The
symbol is used to identify those connections which remain to be done during on site
installation.
−
The
symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.
DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS
MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.
Blue
Yellow
Orange
HV TANK
X−RAY TUBE
CABLE #
TUBE−I (+)
ANODE
ANODE
TUBE−I (−)
CATHODE
CATHODE
X−RAY TUBE HV
CONNECTIONS
1
2128497
ST1−1
ST1−3
ST1−5
ST1−10
ST1−11
Clamped
ST1−7
Black
White
Green
Brown
2
3
J100−10
J100−12
J100−6
(Control Board 3A1)
46−329288G2
4
STATOR
CABLE
XT Suspension Power Supply CABLE
46−195120G4
HV
TERMINALS
X−RAY TUBE
(Maxiray−75/100)
MPH A1 A2 MODULE
(Stator Connector
Panel)
PRINTER−X Power Supply
CABLE
ST1−12
Shield
2104204
XJ1A/B−1
TSxx−L
TSxx−N
GND STUD
MPH 50/65/80 CABINET
PRINTER − X
MPH A9 MODULE
XT SUSPENSION
Black
White
Green
Brown
Blue
Orange
Yellow
Shield
GND Screw
(1)−Pressure Switch
(2)−Pressure Switch
(4)−Terminal Strip
(5)−Terminal Strip
(7)−Terminal Strip
(6)−Terminal Strip
(8)−Terminal Strip
MPH A6 A3 MODULE
(Distribution Board)
XJ1A/B−2
Ground Plug
MPH A1 MODULE
(AC Distribution
Panel)
F15−2
F14−2
Ground Plug
120/240Vac
0Vac
Ground
120/240Vac
0Vac
Ground
J100−5
J100−1
0Vac
240Vac
Gnd
jumper
MIS 10680
W34
MIS10707
L
L
C
S
M
C
S
M
46−195120G4
L
L
C
S
M
C
S
M
MIS 107
0Vac
230Vac
Gnd