Job Card IST 001
3 of
4
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
77
ILLUSTRATION
8
MPH CABINET
MPH A6 A3
Distribution Board
MPH A6 A4
Wall Bucky Board
5.1
Connection of Door Open safety switches (see schematic).
With this feature, X
−
Ray generation is inhibited while the door of the X
−
Ray room remains open.
Connect the safety switch to J4 pin 1 & 2 (or a jumper if there is no safety switch) if Wall Bucky
Board.
5.2
Connection of Room Lamp, X
−
Ray Exposure on Signal Lamp and Bucky.
The generator is configured in the factory to supply Buckies with 220VAC. If the Buckies must be
supplied in 115VAC, change the fuse F9 of the Room Distribution board MPH A6 A3 from position
F9A to position F9B. All the Buckies are now supplied on 110VAC ( it is not possible to mix the
voltages). The room lamps are also supplied in 110VAC.