System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
24
MPH CONFIGURATION CHECKS
CONFIGURATION
Re
−
do MPH necessary Configuration
1 Person
ÍÍ
ÍÍ
A24
See MPH SM 2137354
−
100, Job Cards
LG001, RG001 and RG002
NOTE: Connect Service Terminal as described
in Job Card LG 001
Step
Description
Done
Supporting Documents
ÍÍ
ÍÍ
Mark each box when done
From previous page
CONFIGURATION
D Re
−
do necessary Sub
−
System
Components Configuration
ÍÍ
ÍÍ
See C40E SM 2150259
−
100 (Chapter 2)
SG60 SM 2150258
−
100 (Chapter 2), SG100 SM
2150257
−
100 (Chapter 2), XT SM 46
−
019803 or
2118910
−
100 (section 12), RS 85 SM Mounting
Instructions 0327 7223 (section 3)
A25
ÍÍ
ÍÍ
REMARK:
The Run
−
in & Stability check is
performed at the factory. See Technical
Publications, Dir. 46
−
013804
To redo on site ONLY if necessary (refer to
System Inspection Report Completion date).
Use procedure for MVP generators.
RUN
−
IN & STABILITY CHECK
Rotating
anode Tube
1 Person
1 Hour
A26
1 Person
Continues on next page
Refer to Inspection Report 2179651 TDR or to
Inspection Report 2167647 TDR, or 2172166 TDR
section 1.3
Refer to Inspection Report 2179651 TDR, section
4.3 for 40E; 6.3 for SG; 7.2 for RAFL
−
S
Refer to Inspection Report 2179651 TDR or Inspection
Report 2167647 TDR, or 2172166 TDR section 2.2
Refer to Inspection Report 2167647 TDR, or 2172166 TDR
section 4.3 for C40E; section 5.3 for SG; section 6.2 for
RAFL
−
S
or
Содержание Compax 40CE
Страница 138: ......