Job Card IST 001
4 of
4
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
62
3. Rotate the Tube support arm so the collimator is facing down, then remove the Collimator
rear cover and connect to J/P2 and J/P1 plugs, Collimator cable Refer to MIS MAP and MIS
CHART of this Manual).
Note:
The J/P1 plug is used only with Ultranet Sa Collimator (Auto) when JP2 plug is used with both
SA and SM Collimator.
4. Re
−
install the Collimator rear cover and reconnect small ground wire if removed in Step 3.