Version No.: 3
SYSTEM COMPAX 40E
Job Card IST 001
1 of 8
Purpose:
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
Date: November 10, 2006
Time:
Personnel:
1
Do not remove the shipping anchor rod from the Vertical
carriage until instructed to do so See step A8 in the Steer-
ing Guide. (fig. 1)
Slightly loosen the Tube Stand Arm anchor bolts to insert
the Tube Stand Arm into the brackets.
WARNING
shown without
cabling
Fig.1
Note:
Note:
The Collimator Interface Plate is adjusted
parallel to the X
−
Ray tube axis so that the
middle of the two focal spots is centered on
the Collimator Alignment Tool crosshair.
During the HLA procedure in factory:
−
The Tubestand arm is installed, levelled and
adjusted.
−
The X
−
Ray tube is aligned with the Collimator
Interface Plate left which is installed.
−
In the field it is only necessary to mount the
Tube Assy on the Tube Support Arm and
install the collimator with the following
instructions.
DO NOT REMOVE THE ROD
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
67
SECTION 1
SPARES REQUIRED
SECTION 2
EQUIPMENT REQUIRED
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PRELIMINARY STEPS
SECTION 5
INSTALLATION PROCEDURES