6-2
CIC Pro™
2026419-033E
Configuration
Pre-configuration process
Pre-configuration requirements
All the CIC Pro centers connected to the CARESCAPE Network IX and MC
networks must comply with the following configuration requirements:
All devices must have the same time zone settings.
The
Automatically adjust clock for daylight saving changes
check box
must remain
UNCHECKED
at all times.
All MC network IP addresses must be in the same IP scheme with the same
subnet mask.
All IX network IP addresses must be in the same IP scheme with the same subnet
mask.
In the hierarchy of multiple compatible CIC Pro center hardware and software
versions that co-exist, it is important to assign a set of the highest MC network IP
addresses to the highest software version. For example, if you are installing CIC
v5.1.x to an existing CARESCAPE Network comprising of CIC v4.0.7 or later,
you must allocate a set of the highest MC network IP addresses to all of the CIC
Pro center’s hardware running CIC v5.1.x. No lower version CIC Pro center can
have an IP address on the MC network that is greater than the IP address of a
higher version CIC Pro center.
Use the Check Centrals utility to check for the following on all the CIC Pro centers on
the network:
Time zone settings
Daylight saving time (DST) status
CARESCAPE Network IP address errors
Status of 4.1.1-1 patch application on CIC Pro center v4.1.1 (available only with
the release of CIC Pro center v5.0.7 or later)
NOTE
The Check Centrals utility does
not
discover any version of Central Scope.
The Check Centrals utility
will
discover CIC Pro center hardware running
v1.5, but will not report its IP configuration information. This means that
you must physically locate these devices if they exist on the network. Please
read step
on page
The Check Centrals utility will
not
discover any non-CIC Pro center device
(e.g., printers, Aware Gateway, PDS) on the network.
The Check Centrals utility is also supplied separately. (For the part number of the
Check Centrals utility, see
Disaster recovery software kit and Service Tools CD on
.) You can run the utility directly from the CD on BCM and Nightshade
platforms. However, this utility must be copied from the CD to an NTFS formatted
USB memory stick for use on Bedrock platforms (v4.1.1 and greater).
Pre-configuration instructions
1.
Read and understand the following Caution and communicate this information to
the biomedical/clinical staff:
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......