6-26
CIC Pro™
2026419-033E
Configuration
Configure ADU alarm audio setting, if applicable
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
CAUTION
OUT-OF-UNIT ALARMS — If the CIC Pro center is configured to
alarm for out-of-unit alarms, any patient displayed on that out-of-
unit CIC Pro center, and is alarming, can have their alarms silenced
from the out-of-unit CIC Pro center.
CAUTION
AUDIBLE ADU ALARMS MAY NOT SOUND—Depending on
the configuration of your CIC Pro center, audible ADU alarms may
not sound at the CIC Pro center for any in-unit patient beds.
Only on-screen ADU alarm indicators display unless the CIC Pro
center is configured to also sound audible alarms.
Complete the following procedure to configure the
ADU Alarm Audio
setting on the
CIC Pro center:
1.
If you have not already logged onto Webmin,
2.
Click
Configuration > Set Flags
.
NOTE
The
Set Flags Module
screen shows the current set flags settings, which can
be modified.
3.
Change the settings in the
ADU ALARM AUDIO
field as required:
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......