3.
Analyze the Check Centrals output:
a.
Identify time zone or daylight saving time network errors:
Verify that all the CIC Pro centers configured on the CARESCAPE
Network are listed.
NOTE
The target CIC Pro center running the Unity Time Zone Discovery
Tool utility is the baseline device used for determining time zone or
daylight saving time errors.
If errors are identified for either the time zone (TZ) or daylight saving
time (DST) settings, correct the time zone settings as per instructions in
Set the time zone on page 6-34
b.
Identify CARESCAPE Network IX and MC addressing scheme errors:
Verify the CARESCAPE Network IX and MC addressing schemes
match for all the CIC Pro centers on the CARESCAPE Network.
If the addressing schemes do not match, complete Network IP address
configuration for each CIC Pro center that requires IP address changes,
as per instructions in the appropriate CIC Pro Clinical Information
Center Service Manual.
4.
If there is any CIC Pro center hardware running v4.1.1 on the network, make sure
that the v4.1.1-1 patch is applied.
5.
Make sure that
only
the following compatible CIC Pro center versions co-exist in
a given
care area
:
CIC v5.1.x
CIC v5.0.x
CIC v4.1.1-1
CIC v4.0.7 or CIC v4.0.8
NOTE
CIC Pro center hardware running v1.5, or any version of Central Scope, can
exist independently in a separate
care area
, but not in the same care unit
name with the software versions listed above.
6.
If there is one or more Aware Gateway on the network, configure the Time
Master settings at the Aware Gateway. See the latest Aware Gateway Service
Manual for instructions. Consult Tech Support and hospital IT if any assistance is
required.
7.
Investigate if there is any unauthorized, non-GE medical equipment connected to
the MC/IX network. If any unauthorized non-GE equipment is found or reported
to be connected to the CARESCAPE Network, read and understand the
following Warning and communicate this information to the hospital IT/
biomedical or clinical staff before proceeding further:
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......