Configuration
2026419-033E
CIC Pro™
6-69
CAUTION
RESTART AFTER ADVANCING THE TIME SETTING — When
advancing the time 72 hours or greater, you must restart the CIC Pro
center to continue collecting full disclosure data. If you do not
restart the CIC Pro center, full disclosure data collection stops and it
may also cause other data integrity and CIC Pro center performance
issues.
All the CIC Pro centers connected to the CARESCAPE Network IX and MC
networks must all use the same time zone setting. Before making any changes to the
time zone, time-of-day, or the date settings, you must make sure the time zone and
Daylight Saving time (DST) settings match for all of the CIC Pro centers on the
CARESCAPE Network.
NOTE
To change the time zone, daylight saving time, time-of-day, or the date on a CIC
Pro center using v4.0.x or v4.1.1 software or later, see the CIC Pro Clinical
Information Center Service Manual that was provided with your equipment. Use
the Check Centrals utility.
See Pre-configuration instructions on page 6-2.
This section describes the procedures required to set the time and date of a CIC Pro
center using v5.1.x software or later.
1.
If you need to change the time zone or daylight saving time (DST) settings, be
sure to change the time zone and DST settings before you change the time-of-day
or the date settings.
2.
If you have not already logged onto Webmin,
3.
Click
Configuration > Time Date
.
4.
On the
Time Date
window, make the applicable selections:
a.
In the
Date
field, choose the date from the drop-down list.
b.
In the
Month
field, choose the month from the drop-down list.
c.
In the
Year
field, choose the year from the drop-down list.
d.
In the
Time
field, choose the appropriate time-of-day (hours:
minutes:seconds) from the drop-down lists.
5.
Click
Change Time
to update your changes.
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......