Checkout procedures
2026419-033E
CIC Pro™
7-7
Check operation of audible alarm tones
WARNING
The CIC Pro center audible alarms will not sound for bedside
patients configured to the
Operating Room
mode.
NOTE
All clinical setting values must be received from the GE Clinical Application
Specialist (CAS) or from the nursing director of the care unit. Consult with your
CAS or hospital staff about the use of these functions and if they should be used
with your configuration.
NOTE
Active alarms override this test.
1.
From the single patient viewer menu, click
Monitor Setup
>
Alarm Control
.
2.
Click
Alarm Help
>
Advisory Alarm
.
3.
Verify the alarm sounds through both speakers.
Check access to all other care units
1.
From the multi-patient viewer menu, click
View Other
.
2.
Verify that the list shows all care units that should be networked.
Check the operation of the Citrix application viewed at the CIC Pro
center
It is the responsibility of the hospital’s IT or Biomedical department to meet the
required CIC Pro center configuration needs to check this functionality.
Check the hospital intranet browser functionality
It is the responsibility of the hospital’s IT or Biomedical department to meet the
required CIC Pro center configuration needs to check this functionality.
Check the status of locked beds
Based on the needs of the hospital staff, ensure that the required beds are locked.
Set up locked beds on page 6-53.
Check the language settings
In the multi-patient viewer window, check if the patient names are displayed the
appropriate language.
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......