8-20
CIC Pro™
2026419-033E
Troubleshooting
Log files
Download log files
When you contact GE Service, you may have to provide required log file information.
1.
If you have not already logged onto Webmin,
2.
Click
Diagnostics > Logfiles
.
Hard Disk temperature out of range (X - X). Currently: X °C.
CPU temperature out of range (X - X). Currently: X °C.
Chassis Fan 1 speed out of range (X - X). Currently: X rpm.
(MP100 platform)
Fan A speed out of range (X - X). Currently: X rpm.
(Legacy platform)
Chassis Fan 2 speed out of range (X - X). Currently: X rpm.
(MP100 platform)
Fan B speed out of range (X - X). Currently: X rpm.
(Legacy platform)
Internal speaker is unplugged.
External speaker is unplugged.
Hard Disk Drive failure.
CAUTION
LIMITED DATA STORE IN FULL DISCLOSURE —If rotating the hard drive fails, the
system can be used for patient monitoring, but with limitations on the amount of data
stored in full disclosure, and provided that the hard drive be replaced within five days of
hard drive failure notice on the CIC Pro center.
Flash Disk Drive failure.
(MP100 platform)
Warning! System resources are running low. A Preventive Maintenance action is
required.
1
Warning! Available system resources are running low. System restart is required
to correct the problem. Patients will not be monitored at this Central while the
System is restarting. If the system is not restarted now, it will restart automatically
in approximately X minutes.
2
1
The system resource indicator changes to yellow until this condition is resolved.
2
The system resource indicator changes to red until this condition is resolved.
Message
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......