Configuration
2026419-033E
CIC Pro™
6-39
Alarm Volume
WARNING
ALARMS — Do NOT rely exclusively on the audible alarm system for Bedside Monitoring. Adjustment of CIC Pro
center alarm volume to a low level or OFF during Bedside Monitoring may result in inability to hear the alarm and
a hazard to the patient. Remember that the most reliable method of Bedside Monitoring combines close personal
surveillance with correct operation of monitoring equipment.
After connecting the monitor to the central station and/or nurse-alert, verify the function of the alarm system.
Repeat this verification periodically, including a check of all connected speakers.
NOTE
All clinical setting values must be received from the GE Clinical Application Specialist (CAS) or from the
nursing director of the care unit. Consult with your CAS or hospital staff about the use of these functions and
if they should be used with your configuration.
Set the alarm volume for the CIC Pro center.
NOTE
The
Minimum
alarm volume can only be set via the Service mode. You can set the
Current
alarm volume to
any level above the
Minimum
via the user or Service mode.
All clinical setting values must be received from the GE Clinical Application Specialist (CAS) or from the
nursing director of the care unit. Consult with your CAS or hospital staff about the use of these functions and
if they should be used with your configuration.
Current
: Choose any value from the
Minimum
alarm volume setting or above. This
Current
alarm volume
setting determines the actual alarm volume.
Minimum
: Choose the minimum alarm volume level. The minimum alarm volume in user mode cannot be
adjusted below this set value.
Real-Time Trend
Graph
Configuration
Enable or disable the display of real-time trend windows in the multi-patient viewer. The real-time trend window
displays the recent patient trends for a maximum of two parameters in a graph format. You can configure the Real-
Time Graph for each patient while in user mode only.
To enable the display of the real-time trend windows in the multi-patient viewer, click the
Display Real-Time
Trend Graph
checkbox.
Option
Function
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......