Un-interruptible power supply (UPS)
WARNING
Connect the UPS to the CIC Pro center(s) and display monitor(s)
only. Do not connect printers or other devices to a UPS, as such
devices may shorten estimated run-times. If AC line power is not
restored before UPS run-time is exceeded, the CIC Pro center
improperly shuts down and patients will not be monitored.
WARNING
Without a UPS, power line outages may result in:
Improper shutdown of the CIC Pro center, causing lengthy disk
scan procedures on reboot.
Data loss.
If power to the CIC Pro center is lost, patient monitoring
information will no longer be displayed or stored.
GE recommends using a UPS with the CIC Pro center. See
for a listing of UPS units available for the CIC Pro center. Follow the manufacturer’s
recommendations for installing the UPS.
NOTE
The CIC Pro center draws approximately 40 watts of power. The displays each
draw approximately 50 watts of power. See the table on page
The following table identifies the estimated wattage for CIC Pro center MP100
systems and 19-inch medical displays:
The actual UPS run-times are variable, and are affected by the following conditions:
Battery age
Ambient temperature
Site-specific UPS usage patterns
Load characteristics
Number of CIC
Pro center
MP100
systems
Number of CIC Pro center
19-inch medical displays
Total load with CIC Pro center MP100
system(s) and CIC Pro center 19-inch
medical display(s) (estimated watts)
One
One display (primary)
75 W
One
Two displays (dual/
secondary)
115 W
Two
Four displays (Two primary,
Two secondary)
230 W
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......