4-10
ApexPro™
2001989-200E
System setup
Patient age
WARNING
INCORRECT ALGORITHMS, ARRHYTHMIA PROCESSING
AND CALCULATIONS BASED ON PATIENT AGE — After
manually updating or automatically retrieving patient demographic
information from a network database,
always
confirm that the
entered patient’s date of birth matches the patient’s actual date of
birth. Otherwise the appropriate age-related algorithms, arrhythmia
detection, and calculations will not be applied.
Detect Pace
Enable or disable pacer detection. Choices are:
Pace 1
,
Pace 2
, and
Off
.
Off
is the default.
NOTE
Selecting
Off
turns pacemaker detection off. It does
not
perform pacemaker detection.
Pace 1
or
Pace 2
must
be used with pacemaker patients.
pacemaker patients on page 7-13.
NOTE
Temporary changes may be made for a specific patient
via the single patient viewer >
Monitor Setup
>
ECG
.
Patient Age
Set
Patient Age
. Choices are:
0-2 Years
,
3-11 Years
,
11-13
Years
, and
Adult
NOTE
Temporary changes may be made for a specific patient
via the single patient viewer >
Admit
.
Transmitter Alarm Pause
Enable or disable transmitter alarm pausing. Choices are:
Enable, Disable
or
Off
.
This option sets the unit default for all telemetry patients
admitted to the CIC Pro center.
NOTE
Temporary changes may be made for a specific patient
via the single patient viewer >
Alarm Control.
Alarm Pause
Breakthrough
Turn on or off
Transmitter Alarm Pause
breakthrough.
Choices are:
Always On
or
Always Off
.
Always On
is the
default.
This option sets the unit default for all telemetry patients
admitted to the CIC Pro center. It cannot be changed on an
individual patient basis.
Event Marker
(ApexPro CH transmitter only) Turn on or off
Event Marker
alert.
Off
is the default.
1
Barbara J. Drew, RN, PhD, FAAN (2000).
Value of Monitoring a Second Precordial Lead for
Patients in a Telemetry Unit
, GE Medical Systems (order document number M04243ME0)
Option
Function
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......