D-4
ApexPro™
2001989-200E
Troubleshooting
369B, 374B
The right arm electrode is moved down to the fifth intercostal space, and the left leg
electrode is moved up to the fifth intercostal space.
NOTE
After all electrodes are in place, ensure that a minimum of 1/2 mV of signal is
present on each lead (I, II, III, V).
Interface connector ports
Pace detection performance is optimized with proper lead application and correct use
of the serial interface connector ports. If you are experiencing degraded pace
detection performance, verify that all leads are properly attached to the patient and
verify that any connected serial device(s) are in the appropriate serial interface
connector ports.
The inside port, labeled
2
on the dust cover, is for use with continuous
monitoring serial devices, such as SpO2.
The outside port, labeled
1
on the dust cover, is for use with episodic monitoring
serial devices, such as NBP.
Problem: Inaccurate pacemaker detection
Solution:
Use pacemaker processing:
1.
Click on the patient’s
ECG
tab.
2.
In the
Detect Pace
section, select either
Pace 1
or
Pace 2
.
Solution
: Exchange the right and left arm leads and perform a relearn.
Notes
In general, BE AWARE that a pacemaker pulse could be falsely counted as a
QRS during asystole.
Pace 1
mode analyzes the presence of a pacer spike, assesses the waveform for
residual pacemaker energy, and determines the presence of an R wave following
the pacer spike. If an event occurs during the first few milliseconds following the
pacer spike, it will be counted.
RA
LA
V1
LL
RL
L
C1
R
F
N
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......