5-4
ApexPro™
2001989-200E
Alarms
When configured, a graph prints automatically when a patient experiences a
Crisis
or
Warning
alarm. Arrhythmia alarm graphs run until the end of the alarm event or
manually stopped by the user. The printer prints the 10 seconds of data that occurred
immediately before the event, and prints for the duration of the event. The printer
stops printing when the patient returns to a normal rhythm. If a printer is not available
at the time of the alarm event, a 20-second graph is saved. This saved graph will print
when a printer becomes available.
You can temporarily adjust patient status alarm levels and limits.
System status alarms
System status alarms are triggered by network or equipment problems. They are of
lesser priority than patient status alarms.
There are three severity levels of system status alarms:
System Warning
: Serious network or equipment problems.
System Advisory
: Network or equipment problems.
System Message
: Additional information only.
The CIC Pro center’s response to system status alarms is as follows.
Managing patient alarms
To support a patient’s unique arrhythmia or parameter condition, you can temporarily
adjust a patient’s
Parameter Limits And Alarm Levels
The bedside monitors automatically stores a 10-second strip for all
Advisory
,
Warning
, or
Crisis
arrhythmia events. The CIC Pro center can retrieve and display
the 10-second strip stored at the bedside monitor.
3
For telemetry patients only, the factory default for this setting is
Always on
. To change this
setting, see the
Telemetry Unit Defaults
in the appropriate CIC Pro center service manual.
Indicator
System Warning
System Advisory
System Message
Alarm tone
Repeating foghorn
Single foghorn
No
On-screen message
Yes
Yes
Yes
Colored patient
window border
1
2
1
When using the multi-patient viewer, the patient window border briefly flashes on and off at the
start of the alarm and then stays on until the alarm is silenced or the alarming condition
ends.
2
Only patient status
Crisis
and
Warning
alarms and system status
Warning
alarms activate the
colored border. Patients selected for single patient view have a white border in the multi-
patient viewer.
Yellow
No
No
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......