Troubleshooting
2001989-200E
ApexPro™
D-9
Graph messages
The following messages appear in the patient’s bed window on the CIC Pro center
display. These relate to running graphs at the printer.
ARTIFACT ARR SUSPEND
All artifact begins at level 1. Sustained artifact progresses to
level 2 when noise on ECG lasts for 20 of the last 30
seconds.
Level 1 — Upon immediate detection of artifact, the
message
ARTIFACT
is displayed. There is no alarm
tone. Only lethal arrhythmia processing is available.
Level 2 — Arrhythmia monitoring is suspended in this
condition. A system warning alarm sounds, no
arrhythmia messages are displayed, no graph strips run
and no arrhythmia events are stored. Heart rate and
PVC values change to X, an additional message,
ARR
SUSPEND
, is displayed.
LEADS FAIL
All leads have failed, right leg lead failed, lead wires
unplugged or reference lead failed. If set to
Warning
or
Advisory
level, a system alarm is heard. This will self-
silence if condition is corrected, or the user can silence for
one minute with the
Silence Alarms
button at the bottom of
the CIC Pro center display.
NO TELEM
The
NO TELEM
alarm occurs in two situations:
1.
The patient moves out of range.
If the transmitter is out of range for more than 30
seconds, the
NO TELEM
message displays. Should a
LEADS FAIL
condition occur prior to a
NO TELEM
condition, the
LEADS FAIL
condition takes priority. The
LEADS FAIL
message is displayed along with the
NO
TELEM
message.
2.
Transmitter batteries are extremely low or dead.
If batteries are extremely low or completely dead, the
NO TELEM
message appears, a
CHANGE BATTERY
message appears in the patient’s bed window, and the
audible alarm sounds. Activating the
ALL ALARMS
OFF
feature prevents the audible alarm from sounding.
OFF NETWORK
In the
Combo
mode, telemetry patient data is provided from
the bedside monitor to the CIC Pro center. If the bedside
monitor is removed from the network or becomes otherwise
unavailable, the
NO COMM
message displays first on the
CIC Pro center and there is a loss of monitoring for about 45
seconds, then waveforms return and the
OFF NETWORK
message is displayed below the HR parameter window.
The CIC Pro center will then display patient data directly from
the telemetry bed along with this message in that CIC Pro
center bed window. Should the bedside monitor reappear on
the network, the
Combo
monitoring application will
automatically resume and the alarm will be cleared.
This alarm is also generated if the receiver system is
removed from the network or becomes otherwise
unavailable. The bedside monitor does not sound alarms, but
the alarms must be turned back on at the bedside monitor if
they were previously paused or off.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......