7-28
ApexPro™
2001989-200E
Viewing real-time patient data
At the CIC Pro center, the signal strength indicator consists of 0, 1, 2, or 3 (strongest)
asterisks, depending on the strength of the signal.
Proper environmental conditions and probe attachment help ensure a strong signal.
WARNING
In the monitoring of patients the coincidence of adverse conditions
may lead to a disturbed signal going unnoticed. In this situation
artifacts are capable of simulating a plausible parameter reading, so
that the monitor fails to sound an alarm. In order to ensure reliable
patient monitoring, the proper application of the probe and the signal
quality must be checked at regular intervals.
Error messages
If the probe is not correctly attached to the patient and data is not verifiable, one of the
following error messages may appear in the patient’s bed window at the CIC Pro
center:
Oximeter device
SPO2 PROBE OFF
SPO2 PROBE
Dinamap Pro monitors
SPO2 NO DATA
SPO2 PROBE
If any of the listed messages appear, check the position of the probe or replace the
probe.
See SpO2 messages on page D-5.
If the problem persists, call GE Service or
contact your sales/service representative.
Pressures
The Pressures tab sheet allows you to view and modify settings specific to the viewed
telemetry patient’s NBP display. Settings may be viewed for any patient. However,
you can only modify settings for patients who are admitted to a bed in your unit.
NOTE
The NBP tab is labeled
Pressures
because other invasive pressures settings are
available on this tab sheet for bedside monitored patients only.
Invasive pressure monitoring is not an option for telemetry patients. Therefore,
only NBP information appears on this tab sheet when monitoring a telemetry
patient.
Non-invasive blood pressure control settings
NOTE
A non-invasive blood pressure device must be connected to transmitter to
configure settings.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......