D-8
ApexPro™
2001989-200E
Troubleshooting
Messages
Alarm messages
The following messages appear in the patient’s bed window at the CIC Pro center.
CHANGE BATTERY
An NBP measurement was attempted with low batteries. Change the batteries in
the blood pressure monitor and try another measurement.
FAIL
A hardware failure has been detected in the blood pressure monitor. In the U.S.,
contact GE Service. Outside the U.S., contact your local sales/service
representative.
LEAK
The NBP cuff is loose or there is an air leak in the cuff or tubing. Check that the
cuff is on snugly. Check the connection between the cuff and the tubing. Check
the connection between the tubing and the blood pressure monitor. Try another
measurement. If the problem persists, contact your local sales/service
representative.
LOW INFLATION PRESS
This message appears when K-sounds are detected immediately upon inflation.
The inflation pressure is too low for proper NBP measurement. Try another
measurement or adjust dynamic/fixed inflate. If the problem persists, contact your
local sales/service representative.
MOVEMENT
This message appears when there is excessive patient arm movement, or if the
patient’s arm was bent during the measurement. Check the patient and try
another measurement.
SENSOR?
This message appears when the K-sounds on the measurement were too weak,
or not enough sounds were detected. Reposition the microphone and try another
measurement. This message can also appear if the deflate rate is not properly
adjusted. If the problem persists, contact your local sales/service representative.
WEAK PULSE
The patient’s heart rate is erratic. Check that the microphone cable is plugged
firmly into the patient cable, and check that the microphone is positioned correctly
on the patient. Try another measurement. If the problem persists, it could indicate
a defective microphone, microphone cable, or patient cable. Contact your local
sales/service representative.
ALARM PAUSE
All alarms for this patient have been turned off for five
minutes. This is initiated from the transmitter by pressing
both the
Verify Leads
and
Graph
buttons simultaneously.
See Pausing alarms at the transmitter on page 5-5.
ALL ALARMS OFF
All alarms for this patient have been turned off. No graph
strips run, arrhythmia events are not stored, and no audible
tones sound if an alarm condition should occur.
ARR OFF
The arrhythmia program for a selected patient has been
turned off. No arrhythmia messages are displayed,
arrhythmia event s are not stored, no graph strips run, and no
audible tones sound if an arrhythmia alarm condition occurs.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......