5-14
ApexPro™
2001989-200E
Alarms
Parameter alarm levels
To make a change in the telemetry unit default alarm level for a parameter, first access
the
Telemetry Alarm Control Defaults
tab as described in the
Telemetry Alarm
Control Defaults
section in this chapter. Then follow the procedure below.
1.
In
Telemetry Alarm Control Defaults
tab, use the mouse to click in the
Level
field of the parameter for which the alarm level is to be changed. A down arrow
button appears in the field.
2.
Click on the down arrow button. A popup list of alarm level selections appears.
331A
3.
Click on the desired alarm level to select it.
4.
Once you have set the desired level, click the
Apply
button for the changes to
take effect.
5.
If you are finished making changes to the
Telemetry Alarm Control Defaults
tab
sheet, click the
OK
button.
Arrhythmia alarm levels
To make a change in the telemetry unit default alarm level for arrhythmia alarms, first
access the
Telemetry Alarm Control Defaults
tab as described in the
Telemetry
Alarm Control Defaults
section in this chapter. Then follow the procedure below.
1.
In the
Telemetry Alarm Control Defaults
tab, the
Arrhythmia Alarm Levels
for
which unit defaults can be set appears on the right side of the window:
333D
2.
Click in the
Level
field of the arrhythmia alarm you wish to modify. A down
arrow button appears in the field.
NOTE
The
Arrhythmia Alarm Levels
for
ASYSTOLE
and
VFIB/VTAC
cannot be
changed. Therefore, the text in the
Level
field for these alarms always
appears dimmed.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......