Printing
2001989-200E
ApexPro™
9-5
manual graph will be saved in alarm history. Additional data maybe available with
Full Disclosure.
Graph all patients
Clicking on the
button on the CIC Pro center display sends a Graph All Patients
request to all beds displayed on the CIC Pro center, initiating a 10-second graph for
all telemetry and
Combo
patients and a 20-second graph for all bedside patients.
When this option is selected for telemetry patients, graph requests always print at a
speed of 25 millimeters per second.
NOTE
This Graph All Patients function is only available when the single patient viewer
is closed. If a single patient viewer is open, selecting
from the main menu
initiates a printout of whichever tab sheet is in front.
The print process stops automatically. If the
Graph Stop
control key is pressed on the
external DDW, the current patient’s graph stops and the writer begins to print a graph
for the next patient.
If a patient’s data is currently printing or is being saved to print when a Graph All
Patients request is initiated, this patient’s data will not be included in the Graph All
Patients graph. This patient’s data will print independently of the Graph All Patients
graph.
Clicking on the ECG parameter window for a patient whose data is saving will cancel
the Graph All Patients request for that patient.
If, while a Graph All Patients request is in progress, an arrhythmia alarm occurs for a
patient, the alarm data will replace the data that was saved for the Graph All Patients
request. The alarm data will then appear on the graph printout.
If, while a Graph All Patients request is running, a telemetry patient initiates a graph
from his or her transmitter, the saved data for the Graph All Patients graph will be
replaced by data from the patient’s transmitter. The data from the transmitter will then
appear on the graph printout.
Initiating a graph all patients request
To initiate a Graph All Patients request, follow these steps.
1.
Click on the
button on the CIC Pro center display. The Graph All Patients
window opens.
2.
Click on
Limits
or
Waveforms
.
Selecting
Limits
graphs all patient limits.
Selecting
Waveforms
graphs all patient waveforms.
3.
Click on the
OK
button to complete the Graph All Patients request.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......