6-20
ApexPro™
2001989-200E
Managing patients
The
Alarm Pause
feature (if available on your transmitter) is honored at the
monitor.
You should not turn off the monitor until you have discharged the patient
from the monitor.
Second V lead data is not sent to the monitor from the transmitter. If you
wish to see telemetry second V lead data, you must view the telemetry
patient.
When switching ECG monitoring from the monitor to telemetry:
Arrhythmia alarm histories from the monitor are merged in the telemetry
system. CD Telemetry-LAN software version 5 or later is required.
If you discharge the monitor, the telemetry
Arrhythmia Alarm Levels
will be
the same as the
Arrhythmia Alarm Levels
supported by the monitor.
Therefore, when the monitor uses the BASIC software package, only lethal
Arrhythmia Alarm Levels
will be detected from telemetry. If the monitor
has the
CARDIAC
software package, full
Arrhythmia Alarm Levels
will be
detected from telemetry.
When switching ECG monitoring from telemetry to the monitor:
Telemetry is automatically discharged and the most recent 36 alarm histories
are transferred to the monitor.
The ECG limits,
Arrhythmia Alarm Levels
and display defaults are recalled
from the monitor defaults.
NOTE
It is not likely that the
Combo
or
Rover
Combo
monitoring modes are used when
the patient-monitor type is
OPERATING ROOM
.
NOTE
Users should be aware of a possible time discrepancy between the waveforms
from the telemetry device and the waveforms from the monitor. Users should not
consider these waveforms to be synchronous. If absolute synchronicity is desired,
Combo
mode should be discontinued and the ECG waveforms should be
acquired via the hard-wired monitor.
NOTE
SpO
2
monitoring with an X-Pod does not occur in combo mode. If SpO
2
monitoring is required while in combo mode, the patient's SpO
2
sensor must be
attached to an SpO
2
device at the monitor.
Constraints
For bedside monitors that allow the ECG parameter to be turned off, the following
constraints apply.
SpO
2
becomes the primary parameter for patient monitoring.
The patient’s heart rate is determined from pulse oximetry.
The SpO
2
and SpO
2
Rate parameter alarm levels become
Warning
.
The SpO
2
pulse search and probe off system status alarm levels become
Warning
.
Connecting the ECG cable to the monitor will
not
automatically turn the ECG
parameter on.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......