Alarms
2001989-200E
ApexPro™
5-15
3.
Click on the down arrow button. A popup list of alarm level selections appears.
4.
Click on your choice to select it.
5.
Once you have set the desired level, click the
Apply
button for the changes to
take effect.
If you are finished making changes to the
Telemetry Alarm Control Defaults
tab
sheet, click the
OK
button.
System alarm levels
WARNING
ADJUSTING SYSTEM ALARM LEVELS — The
Leads Fail
alarm indicates that one or more electrodes are not connected to the
patient and, as a result, there is loss of all waveforms and arrhythmia
analysis. The
ARR SUSPEND
alarm indicates that arrhythmia
conditions are not being detected and therefore alarms associated
with arrhythmias will not occur. The
Leads Fail
and
ARR
SUSPEND
alarms should be adjusted to a lower priority level only
by experienced qualified personnel and with great caution.
Adjusting these alarms to a lower priority level may result in
reduced awareness of conditions that indicate the loss of patient
monitoring.
NOTE
Some alarm conditions may have a higher priority level at the bedside monitor. In
Combo
mode, the alarm levels come from the telemetry settings. For example,
LEADS FAIL
can be set to
Crisis
at a bedside monitor, but the telemetry setting
is a system
Warning
.
To make a change in the telemetry unit default alarm level for system alarms, first
access the
Telemetry Alarm Control Defaults
tab as described in the
Telemetry
Alarm Control Defaults
section in this chapter. Then follow the procedure below.
1.
In the
Telemetry Alarm Control Defaults
tab, the
System Alarm Levels
for
which unit defaults can be set appears on the right side of the window:
430B
2.
Click in the
Level
field of the system alarm you wish to modify. A down arrow
button appears in the field.
3.
Click on the down arrow button. A popup list of alarm level selections appears.
4.
Click on your choice to select it.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......