Viewing real-time patient data
BIGEMINY
Occurs when three or more bigeminal cycles (a ventricular beat followed by
a non-ventricular beat) are detected.
BRADY
Bradycardia is the average of the most recent eight R-to-R intervals at a
heart rate less than the set LOW heart rate limit.
NOTE
The Brady limit matches the low heart rate limit. If the low heart rate
limit is changed, the Brady limit changes.
COUPLET
Occurs when two ventricular beats are detected and have non-ventricular
beats before and after the couplet. The coupling interval must be less than
600 milliseconds.
IRREGULAR
Occurs when six consecutive normal R-to-R intervals vary by 100
milliseconds or more.
PAUSE
Occurs when a 3-second interval without a QRS complex is detected.
NOTE
Some bedside monitors allow the
Pause
time interval to be adjusted.
For more information, refer to the appropriate bedside monitor’s
operator’s manual.
PVC
Isolated premature ventricular complexes occur when a premature
ventricular beat is detected and has non-ventricular beats before and after.
PVC limit
When on, the PVC Limit control displays a PVC counter in the ECG
parameter window. When off, the PVC counter is not displayed. Use the
mouse to turn the PVC Limit control
On
or
Off
. The PVC limits are preset in
Alarm Control
defaults.
R ON T
Occurs when a ventricular complex is detected within the repolarization
period of a non-ventricular beat.
TACHY
Tachycardia is four R-to-R intervals at a heart rate greater than the set HIGH
heart rate limit.
NOTE
The Tachy limit matches the high heart rate limit. If the high heart rate
limit is changed, the Tachy limit changes.
TRIGEMINY
Occurs when three or more trigeminal cycles (a ventricular beat followed by
two non-ventricular beats) are detected.
V BRADY
Adult
—Ventricular bradycardia occurs when a run of three or more
ventricular beats is detected with an average heart rate that is less than or
equal to 50 beats per minute.
0-2, 3-10, and 11-13 years
—Occurs when a run of three or more ventricular
beats is detected with an average heart rate that is less than or equal to 60
beats per minute.
VFIB/ VTAC
Ventricular fibrillation occurs when the ECG waveform indicates a chaotic
ventricular rhythm.
WARNING
VFIB/VTAC
should not be considered a substitute for the
V TACH
arrhythmia call. Efforts to lower the
V TACH
alarm level can result in missed
ventricular tachycardia alarms.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......