Functional and electrical safety checks:
Functional Checkout procedures
2000966-542D
Dash 3000/4000/5000
8-33
Pulse oximetry tests for Nellcor OxiMax SPO2
Equipment
Use the following equipment for these tests:
Nellcor SRC-MAX SpO
2
Simulator (2007650-002).
NOTE
The Nellcor SRC-MAX SPO
2
Simulator is only compatible with
the Nellcor SPO
2
parameter.
Procedures
1. Verify that 2 AA alkaline batteries are installed in the SRC-MAX
Nellcor Pulse Oximetry functional tester.
2. Connect the SRC-MAX Nellcor Pulse Oximetry functional tester to
the 9-pin end of sensor connector from the Dash. Make sure the Dash
is powered.
3. Verify that the IR LED and RED LED drive indicators are both lit on
the SRC-MAX.
4. Verify the SRC-MAX default indicators are set as follows:
HEART RATE = 60 bpm
LIGHT = LOW
%SpO2 = 75
MODULATION = LOW
Allow the Dash a few seconds to obtain a steady reading.
SpO2 Port.
Nellcor SpO
2
cable.
Nellcor compatible
SpO
2
Simulator.
892A
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......