Configuration:
Service menus
2000966-542D
Dash 3000/4000/5000
4-7
MENU SETUP
Provides the following sub-menus: (Refer to
on page 4-11 for detailed procedures.)
ADMIT MENU: STANDARD
Allows you to determine the function of the patient
monitor. The four variables include stationary or
ambulatory (telemetry) patient monitoring with a
monitor that always stays in one room (STANDARD)
or a monitor that moves from room to room (
ROVER
).
MONITOR DEFAULTS PASSWD
Allows you to set the patient monitor so that a
password is Required or Not Required for entry into
the Monitor Defaults menu section. If selected, the
password will be the same as the Service Mode menu
password.
MONITOR SETTINGS
Provides the following sub-menus: (Refer to
on page 4-11 for detailed procedures.)
SET UNIT NAME
Allows changes to the care unit name. After initial
setup, this name should not be changed or
communication to the central station will be corrupted.
Note that the care unit name must be registered
exactly the same in the central station and the patient
monitor.
SET BED NUMBER
Allows changes to the bed number. After initial setup,
this number should not be changed or communication
to the central station will be corrupted. Note that the
bed number must be registered exactly the same in
the central station and the patient monitor.
SET INTERNET ADDRESS
Allows changes to the internet (IP) address. An
incorrect internet address may also prevent the
patient monitor from viewing other monitors on the
network even though the unit names match. Whether
or not this can occur depends on the network topology
at the installed site.
WARNING
Duplication of an internet (IP) address on a network
causes lost data. If you change the factory assigned
internet address, you must first record all
other
internet
addresses used on your network to avoid duplication.
NOTE
An incorrect internet address may also prevent the patient
monitor from viewing other monitors on the network even
though the unit names match. Whether or not this can occur
depends on the network topology at the installed site.
Service Mode menu options (Continued)
Menu option
Description
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......