Troubleshooting:
Battery alarms and messages
2000966-542D
Dash 3000/4000/5000
6-7
Battery alarms and messages
Battery alarms occur when one of the following battery conditions occur:
Low battery
Empty battery
Battery failures
Charger failures
These Battery messages help identify the cause of a battery alarm.
Battery Messages appear in three different locations of the patient
monitor:
ECG waveform area.
Battery Status Information window.
Battery Fuel Gauge icon.
Battery messages displayed in the ECG waveform area
The following battery messages are displayed in the ECG Waveform
Area.
Battery messages displayed in the ECG waveform area
Message
Battery alarm
Cause
BATTERY LOW
System
WARNING
Critical Low Battery
Only 10 minutes per battery of run
time remaining (10 minutes if one battery, 20 minutes if two
batteries).
POWERING DOWN
System
WARNING
Empty Battery
There is no battery run time remaining.
CHECK BATT STATUS
System
MESSAGE
Battery Failure
A minor failure has occurred while using or
charging the battery.
BATTERY ERROR
System
WARNING
Battery Failure
A serious failure has occurred while using
or charging the battery. An incompatible battery may be
installed.
CHECK BATT STATUS
NOTE
INTERNAL CHARGER
FAILED, CALL SERVICE
also appears in the Battery
Status information window.
System
MESSAGE
Charger Failure
Charger communications have failed.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......