Installation:
Dash installation checkout procedure
3-8
Dash 3000/4000/5000
2000966-542D
Dash installation checkout procedure
GE recommends that qualified personnel shall perform the following
tests, and should record the results on the
The Dash unit has been thoroughly tested before leaving manufacturing.
There is no need to re-test the parameters or operation of the unit upon
installation except for the following:
1. Electrical safety test. Refer to
“Patient monitor power-up tests”
3. Battery check. Ensure the battery indicator displays on the patient
monitor without an error.
4. If the Dash is networked:
Connect Dash to the MC network (wired or wireless) with any
other GE bedside monitor.
Select
View Other
.
Select another bed.
Verify you can view another bed over the network.
5. If the Dash has TRAM-rac 2A housing connected via the
Aux
port:
Verify communication with a module in the TRAM-rac 2A
housing by confirming that the parameter for the ICG or SAM
module displays on the Dash.
6. If the Dash has a BISx module connected via the
Aux
port:
Verify communication with the BISx module by confirming that
the parameter for the module displays on the Dash.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......