Functional and electrical safety checks:
Functional Checkout procedures
8-36
Dash 3000/4000/5000
2000966-542D
Procedures
1. Attach the digital manometer, noninvasive blood pressure (NBP)
cuff, tees and tubing, as shown in the illustration below, to the NBP
connector of the patient monitor.
2. Set the digital manometer power switch to the on position.
3. Set the digital manometer range switch to 1000 mmHg.
Using the
Trim Knob
control, access the
Service Mode
menu starting
from the Main
Menu.
1. Select
MORE MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros (e.g. July 4 =
0407).
3. Select
CALIBRATE
>
CALIBRATE NBP
>
CHECK CAL OFF
>
START
.
The text on the menu item changes from
CHECK CAL OFF
to
CHECK CAL IN PROGRESS
.
Verify the readings in the NBP parameter window on the patient
monitor display and readings on the digital manometer are equal (± 1
021B
NBP cuff,
Any size NBP cuff
will be sufficient.
6-inch diameter PVC
pipe (or 1-pound coffee
can):
Wrap the NBP cuff
around this for tests.
NBP tubing
NBP port
Coupling,
Connects manometer
tubing to NBP tubing.
3-way tee,
Connects manometer
tubing, NBP tubing
and NBP cuff tubing.
Digital Manometer with
a range of at least 0 to
1000 mmHg Accuracy
0.5% FS.
Note: A calibrated
mercury manometer
may be substituted.
Manometer tubing,
Approximately two
feet in length.
Coupling,
Connects manometer
tubing to NBP cuff
tubing.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......