Configuration:
Boot Code settings
2000966-542D
Dash 3000/4000/5000
4-19
Set CIC and QS protocol
The CIC and QS Protocol default setting is
Seg50/51
(Segment 50/51)
and should not be changed. However, future products may require
Hilltop
protocol. Use the Boot Code Service Menu to change the CIC
Protocol and QS Protocol.
Set MUSE system protocol
The patient monitor transmits 12SL and ACI-TIPI data over the optional
CARESCAPE Network to the MUSE System and ST Guard. The formats
used for this process are Hilltop and Segment 50/51. Since the Segment
50/51 format does not support 500 Hz ECG data or ACI-TIPI, records
originally stored on the MUSE System in Hilltop format cannot display
at the ST Guard. The default setting is
Hilltop
.
1. Activate Boot Code (
“Boot Code settings” on page 4-18
) and select
SET CONFIGURATION
menu option.
2. Select
MUSE System Protocol
then choose
Hilltop
or
Seg50/51
.
Hilltop
—Sends 12SL records to the MUSE System in Hilltop
format. The 12SL records contain 500 samples per second of ECG
data. ST Guard will
not
be able to retrieve 12SL records stored on
the MUSE System in Hilltop format.
Seg50/51
—Sends 12SL records to the MUSE System in Segment
50/51 format with MAC Rhythm statements. The 12SL records
stored on the MUSE System contain 240 samples per second of
ECG data. ST Guard can retrieve 12SL records from the MUSE
System. ACI-TIPI data can
not
be retrieved at the MUSE System
if this protocol is selected.
3. Select
Exit
and wait for the patient monitor to display the message,
“
Writing settings to Flash config mem...
”
4. Manually reboot the patient monitor to implement this change.
NOTE
Before you can transmit data to the MUSE System and ST
Guard, you must set up the MUSE system’s Location ID and Site
Number.
Do not
configure MUSE system’s location at Site 0. Site 0
appears invalid to MUSE and any message coming from a device
on Site 0 is dumped.
“Service Mode settings” on page 4-11
and select
MONITOR
SETTINGS
>
SET LOCATION
, then select
SET SITE NUMBER
.
Transcutaneous Pace Blank Length
This menu option is reserved for future use and should not be changed.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......