Functional and electrical safety checks:
Functional Checkout procedures
2000966-542D
Dash 3000/4000/5000
8-43
ECG or BP calibration
NOTE
Only perform this test if the ECG or BP out signals from the previous
test are out of specifications.
To complete the ECG or BP calibration, connect a voltmeter to the
patient monitor.
1. Attach the analog output cable (pn 2000633-001) to the patient
monitor.
2. Connect a 5.5 digit precision voltmeter (such as HP34401A, or
equivalent) to the port pin to be calibrated (If ECG: Pin 7 = ECG
signal, Pin 3 = ground. If BP: Pin 6 = BP, Pin 5 = ground.).
3. Access the patient monitor’s
Service Mode.
4. Calibrate the ECG or the BP parameters as follows.
ECG calibration
Using the
Trim Knob
control, access the
Service Mode
menu starting
from the Main Menu.
1. Select
MORE MENUS
>
MONITOR SETUP
>
SERVICE MODE
.
2. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
3. Select
CALIBRATE
>
CAL ECG ANALOG OUT
>
SET ECG LOW
.
4. Adjust the count for –9.0 V ± 5 mV on the meter and press the
Trim
Knob
control.
5. Select
SET ECG HIGH
.
6. Adjust the count for +9.0 V ± 5 mV on the meter and press the
Trim
Knob
control.
7. Select
SET ECG ZERO
.
Adjust the count for 0.0 V ± 5 mV on the meter and press the
Trim
Knob
control.
8. Select
CONFIRM ECG CAL
to confirm or abort the calibration.
BP calibration
Using the
Trim Knob
control, access the
Service Mode
menu starting
from the Main Menu.
022A
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......