Configuration:
Boot Code settings
2000966-542D
Dash 3000/4000/5000
4-21
Chinese Language
NOTE
The following two items are only available when the Chinese
language is selected. The factory default is
disabled
.
Select
3.2s baseline filter coefficent
to enable or disable the 3.2
second baseline filter time constant.
Select
Minimum Waveform Width Reduction
to enable or disable
the size reduction of the LCD waveform line width.
Enable or disable AFIB Identification
NOTE
This feature is already enabled if the Cardiac software option is
purchased.
Select
AFIB Identification
to enable or disable the atrial fibrillation
detection feature.
To verify the option has been enabled on the patient monitor,
1. Select the following menu options:
MORE MENUS
>
MONITOR
SETUP
>
SOFTWARE CONFIGURATION
.
2. Verify that
ATRIAL FIB
is listed under
“
ENABLED SOFTWARE
OPTIONS
.”
To change the status of this feature,
1. Activate Boot Code (
“Boot Code settings” on page 4-18
) and select
SET CONFIGURATION
menu option.
2. Select the
AFIB Identification
menu option.
3. Select
enabled
to activate this option, or
disabled
to deactivate this
option.
4. Select
Exit
and wait for the patient monitor to display the message,
“
Writing settings to Flash config mem...
”
5. Manually reboot the patient monitor to implement this change.
6. Refer to the operator’s manual to configure the patient monitor to
use this feature.
Enable or disable IntelliRate
IntelliRate heart rate calculation feature is enabled from the factory.
To verify the option has been enabled on the patient monitor,
1. Select the following menu options:
MORE MENUS
>
MONITOR
SETUP
>
SOFTWARE CONFIGURATION
.
2. Verify that
ECG INTELLIRATE
is listed under
“
ENABLED
SOFTWARE OPTIONS
.”
To change the status of this feature,
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......