Functional and electrical safety checks:
Functional Checkout procedures
8-52
Dash 3000/4000/5000
2000966-542D
Wireless LAN test (option)
Purpose
These procedures check the patient monitor for wireless communication.
While moving a roll-stand with a wireless patient monitor from edge to
edge of the predetermined wireless network coverage area, check for the
following at the edges of the wireless network:
the
SIGNAL LEVEL (RSSI)
is equal to or greater than the
minimum user-defined value,
the
TRANSMIT RATE
is equal to or greater than 5.5, and
a constant waveform from a stationary wireless/wired patient
monitor displays without any loss of waveform.
NOTE
This is a verification of the patient monitor and must not substitute a
professional site survey, installation and verification of the areas
designed for the wireless network. The connectivity and coverage of
the wireless network is verified through the
SIGNAL LEVEL
(RSSI)
status and
TRANSMIT RATE
status in the wireless patient
monitor.
Equipment
Use the following equipment and information for this test:
Wireless Dash patient monitor being tested.
Roll-stand (optional)
Stationary wireless/wired patient monitor
Multi-parameter patient simulator or an equivalent ECG patient
simulator.
Contact the hospital IT department to obtain the following
information:
access point settings for
SSID
and
SECURITY
, and
wireless network coverage area.
Procedures
1. Connect the patient simulator to the stationary wireless/wired
patient monitor and generate a waveform.
2. Admit the stationary wireless/wired patient monitor to the
CARESCAPE Network MC.
3. Configure the wireless Dash patient monitor to communicate with
the wireless network as follows:
a. On the patient monitor, select
MORE MENU
>
MONITOR
SETUP
>
SERVICE MODE
.
b. Enter the password.
c. Select
MONITOR SETTINGS
>
CONFIGURE WIRELESS
LAN
.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......