Functional and electrical safety checks:
Functional Checkout procedures
8-26
Dash 3000/4000/5000
2000966-542D
Temperature tests
Equipment
700/400 Series dual temperature adapter (2016998-001)
Temperature simulator cable
Temperature test procedures
1. Set the patient simulator temperature output to 37
°
C.
2. Attach the temperature adaptor cable to the
TEMP/CO
connector of
the patient monitor.
3. Set the switch on the temperature adaptor to the 400 or 700 position
depending upon customer preference.
4. Attach the temperature simulator cable from the
SERIES 400 or
700 TEMPERATURE OUTPUT
connector of the patient simulator
to the
T1
connector of the temperature adaptor.
5. Verify a TEMP parameter window appears on the patient monitor
display with a T1 reading of 37.0
°
±0.4
°
C.
6. Move the temperature simulator cable from the
T1
connector of the
temperature adaptor to the
T2
connector of the temperature adaptor.
7. Verify a T2 reading of 37.0
°
±0.4
°
C in the TEMP parameter window
on the patient monitor display.
8. Remove the temperature adaptor and temperature simulator cable
from the patient monitor and patient simulator.
017B
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......