Functional and electrical safety checks:
Functional Checkout procedures
8-30
Dash 3000/4000/5000
2000966-542D
3. Change the patient simulator BP output to 240 mmHg.
4. Observe a reading of 240/240 (240) ± 4 mmHg in the BP parameter
window on the patient monitor display.
5. Change the patient simulator BP output to WAVE or to a 120/80
wave form (simulated BP waveform).
6. Observe a distortion-free BP waveform and a reading of
approximately 120/80 in the BP parameter window on the patient
monitor display.
7. Disconnect the BP simulator cable from the
BP
connector of the
patient monitor.
8. Repeat steps for each enabled
BP
connector.
Pulse oximetry tests for GE Ohmeda SPO2 oximeter
Equipment
Use the following equipment for these tests:
GEMS-IT SpO
2
simulator (408610-001 or equivalent).
NOTE
The GEMS-IT SPO
2
simulator is only compatible with the
GEMS-IT SPO
2
parameter.
GEMS-IT SpO
2
simulator cable (700232-004 or equivalent)
Procedure
1. Set the pulse oximetry (SpO
2
) simulator power switch to the off
position.
2. Connect the Nellcor-style SpO
2
simulator cable between the SpO
2
connector of the patient monitor and the SpO
2
simulator.
020A
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......