Functional and electrical safety checks:
Functional Checkout procedures
8-32
Dash 3000/4000/5000
2000966-542D
Pulse oximetry tests for Masimo SET SPO2
Equipment
Use the following equipment for these tests:
Masimo SpO
2
Test Kit (2021087-001). Includes Masimo Tester and
SPO2 Sensor Adapter Cable.
NOTE
The Masimo Tester is intended for use only with Masimo
instruments or instruments containing Masimo SET oximetry
with standard Masimo PC Series patient cable connectors. See
the Masimo “Directions for Use” that came with the Tester for
more information
The PPR and SPO2 values for the Masimo Tester must be within
these limits: Peripheral pulse rate: 61 bpm
1 bpm; and SPO2
value: 81%
3%.
Procedures
1. Connect the Masimo-style SPO
2
simulator cable between the SPO2
connector of the patient monitor and the SPO
2
simulator.
2. Turn on the patient monitor.
3. Verify the following are displayed at the patient monitor: (It might be
necessary to turn the SPO
2
parameter on.)
A waveform with an SpO
2
label.
An SPO
2
% reading between
78 - 84
%.
A PPR reading between
60
and
62
beats per minute.
4. Disconnect the simulator cable from the module.
824C
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......