Configuration:
Boot Code settings
4-18
Dash 3000/4000/5000
2000966-542D
Boot Code settings
To activate the Boot Code,
1. With power on, hold down
NBP Go/Stop
(or
NBP Auto
on Dash
5000) and
ZERO ALL
.
2. Press and release the
Trim Knob
control.
3.
Keep holding
NBP Go/Stop
(or
NBP Auto
) and
ZERO ALL
until
the Boot Code information appears on the display.
Set Defib Sync Voltage and pulse width
The patient monitor controls the analog out signal used to trigger a
defibrillator. Refer to the defibrillator manufacturer’s manual for the
required pulse amplitude and duration.
Configure or change the Marker Out signal of the
DEFIB SYNC
connector as follows.
1. Activate Boot Code (
“Boot Code settings” on page 4-18
) and select
SET CONFIGURATION
menu option.
2. In the
CONFIGURATION MENU
, select:
1 Defib Sync Voltage
:
and choose
5V
or
12V
amplitude.
2 Defib Sync Pulse Width
:
and choose
10 ms
or
100 ms
for
pulse duration.
3. Select
Exit
and wait for the patient monitor to display the message,
“
Writing settings to Flash config mem...
”
4. Manually reboot the patient monitor to implement this change.
Set Line Frequency
Configure or change the patient monitor line frequency to 50 or 60 Hz as
follows. The default is 60 Hz.
1. Activate Boot Code (
“Boot Code settings” on page 4-18
) and select
SET CONFIGURATION
menu option.
2. Select
Line Frequency
then choose
50 Hz
or
60 Hz
line frequency.
3. Select
Exit
and wait for the patient monitor to display the message,
“
Writing settings to Flash config mem...
”
4. Manually reboot the patient monitor to implement this change.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......