Functional and electrical safety checks:
Functional Checkout procedures
8-38
Dash 3000/4000/5000
2000966-542D
Gain calibration test
NOTE
To proceed with the Gain Calibration Test, set up the patient
monitor and test equipment following the guidelines in the NBP
Checkout Procedure,
“Functional Checkout procedures”
1. Connect a cuff and manometer to the patient monitor.
2. Turn the digital manometer on and adjust the range switch to 1000
mmHg.
3. Select
CAL GAIN OFF
>
CAL GAIN OFF
>
START
.
The second line of text on the Cal Gain menu item changes from Cal
Gain Holding to Cal Gain Inflating. The patient monitor starts
pumping up the pressure bulb or cuff—the audible whirring sound of
the NBP pump motors occurs and an increase in displayed pressures
on both the patient monitor and the manometer can be observed.
The pump shuts off at about 250 mmHg, and the pressure drops
slowly to about 240 mmHg before stabilizing. The second line of text
on the Cal Gain menu item changes from Inflating back to Holding.
4. If the pressure continues to drop at a rate of 1 mmHg or more for
every five seconds, there is a leak in the NBP plumbing. If there is a
leak in the NBP plumbing, repair it and restart this calibration
procedure.
5. Select
ENTER CAL PRESSURE
and use the
Trim Knob
control to
select a pressure value that is 1 mmHg lower than the current
manometer reading. When the manometer falls to exactly the value
that you selected in the pop-up window, press the
Trim Knob
control to enter the value.
6. Select
CHECK CAL OFF
>
START
.
7. The text on the menu item changes from
Check Cal Off
to
Check
Cal In Progress
. Verify the pressure readings (shown as
Cuff
in the
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......