Field replaceable units:
Replace main unit parts
2000966-542D
Dash 3000/4000/5000
7-37
Replace DAS assembly
WARNING
DAS ASSEMBLY — Do
not
open the DAS assembly as
this breaks the isolation barrier which may result in
patient death or serious injury. The DAS assembly is a
field replaceable unit only. There are NO field repairs or
adjustments for the DAS assembly.
CAUTION
The patient monitor requires software version 5 or later
to operate with the Masimo SET SPO2 and 4 BP
SuperStat DAS upgrade. If the installation of software
version 5 or later is required, GE recommends that you
FIRST install the software. Then, install the Masimo
SET SPO2 and 4 BP SuperStat hardware upgrade.
1. Remove the two screws holding the handle to the patient monitor.
Refer to page
2. Remove the display assembly from the main unit. Refer to page
3. Remove the four screws anchoring the Data Acquisition System
(DAS) assembly in place.
CAUTION
CAREFULLY remove the DAS assembly so that you do
NOT
hit the components on the CPU/battery housing
assembly.
On the Dash 5000 patient monitor the top two screws attach to a
bracket, not the rear housing as shown.
545A
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......