Functional and electrical safety checks:
Electrical safety tests
8-4
Dash 3000/4000/5000
2000966-542D
Electrical safety tests
General
Electrical safety tests provide a method of determining if potential
electrical health hazards to the patient or operator of the device exist.
Recommendations
Qualified personnel must perform all safety tests presented in this
chapter:
Upon receipt of the device (patient monitor and its associated
equipment). Refer to
“Dash installation checkout procedure”
page 3-8 for more information.
Every 12 months thereafter (Preventive Maintenance). Refer to
on page 5-2 for more information.
Each time the main enclosure is disassembled or a circuit board is
removed, tested, repaired, or replaced (Corrective Maintenance).
Refer to
on page 7-52 for more information.
GE recommends that the qualified personnel performing the tests should
record the values of each required electrical safety test in the
These instructions are intended for every component in the system.
Test equipment
The recommended test equipment required to perform electrical safety
tests is listed below.
Perform electrical safety tests using an electrical safety analyzer per IEC
60601-1, UL 60601-1, EN 60601-1 or CSA C22.2 No. 601.1. The
schematics in the section provide a general understanding of the test
equipment. Actual configuration of test equipment may vary.
The patient monitor being tested should be placed on an insulating
surface.
Item
Specification
Leakage Current Tester
Equivalent to the circuits shown
Digital Multimeter (DMM) (optional based on
leakage tester used and locality)
AC volts, ohms
Ground Bond Tester
0 – 1 ohm
Safety Test Body Kit
1
PN M1155870 or equivalent
1
Instead of the test bodies in the safety test body kit, other applicable test bodies with pins
connected together may be used.
Содержание 001C 051D 003A
Страница 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Страница 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Страница 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Страница 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Страница 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Страница 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Страница 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Страница 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Страница 116: ...Preventive maintenance Clear the stored patient data memory 5 18 Dash 3000 4000 5000 2000966 542D ...
Страница 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Страница 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Страница 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Страница 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Страница 256: ...Functional and electrical safety checks Checkout procedures completion 8 56 Dash 3000 4000 5000 2000966 542D ...
Страница 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Страница 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Страница 273: ...2000966 542D Dash 3000 4000 5000 C 1 C Network disclosure to facilitate network risk management ...
Страница 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Страница 281: ......