TROUBLESHOOTING: Power Source Tests
Revision A
DASH 2000 Patient Monitor
4-7
2000 412-001
Power Cord and Plug
Verify the power cord being used with the monitor is good. The following
are a couple of things to check for in this regard:
¥
Failure of the power cord strain relief is very common. Often times
users of the equipment pull on the power cord itself, rather than the
power cord plug, to unplug the monitor from a wall receptacle. If in
doubt, test for continuity through each conductor of the power cord
connector and plug.
¥
Verify line, neutral, and ground conductors are properly connected to
the power cord plug and are not short-circuited. Rewire and tighten
these, or replace the power cord, as necessary.
Ground Continuity
and Impedance of
Protective Earth
Connection
¥
Verify correct Ground Continuity and Impedance of Protective Earth
Connection as described under ÒPreventive MaintenanceÓ.